Mr. Yuvraj Kunwar is a Certified Sleep & EEG technician (Dept. of Pulmonology & Neurology) in Max healthcare Institute Ltd. He has done Bachelor in Lab Technology & Master in Microbiolog. He started his carrier in Sleep field in 2011 under direct supervision of Dr. Manjari Tripathi (Professor of Neurology) AIIMS New Delhi and then HCL Healthcare(affilated with John Hopkins International medicine).
He has done various type of sleep studies (Diagnostic, Titration, split night, MSLT, Video EEG). He is a Certified Sleep Technician by ISSR (World Sleep Federation Program). He has attended Sleep Seminars, Confrences, workshops, and training & health camps.
He is very happy to be working in this field because this field is new in India. His strengths are self-motivation & dedication towards his work & a good learner.
- Clean using mild soap and water. Distilled water is recommended to avoid damage to the surface of the electrode.
- Scrub lightly using soft bristled brush.
- Disinfect with a sterilization chemical such as Control III for 10-15 minutes per manufacturer’s recommendations.
- Rinse and air-dry.
- Wipe with a cleaning agent that is non-corrosive to plastic.
- Allow assembly to completely air dry before using on a patient.
- Gas sterilization is permitted as needed.
- Wipe with a cleaning agent that is non-corrosive to plastic.
- Allow assembly to completely air dry before using on a patient.
- Gas sterilization is permitted as needed.
- Unplug device from the wall outlet to avoid electrical shock.
- Wipe the outer chassis with a cloth dampened with mild detergent.
- Allow device to dry completely before plugging into electrical wall outlet.
- Do not immerse in water.
- Inspect all parts and replace broken or damaged pieces before putting into service.
- For each individual mask, see package insert for specific cleaning requirements, which vary by mask and manufacturer.
- Pasteurization or other high-level disinfecting processes may be used.
- Some mask discoloring may occur
- Consider using a ding timer as a reminder to remove the mask from disinfectant to prolong its life.
- Wash the head strap daily with warm, soapy water such as Ivory or baby shampoo.
- Rinse thoroughly.
- Air dry.
NOTE: Avoid vinegar, bleach, chlorine or alcohol solutions. Also avoid solutions containing moisturizers, antibacterial agents or scented oils. These products may reduce the life of the head strap.
- Clean tubing with a soft bristle brush while soaking in detergent.
- Rinse thoroughly.
- Soak in an ortho-phthalaldehyde solution such as Cidex® OPA.
- Rinse thoroughly.
- Air dry.
- Store in a dry, dust-free environment
- Cannula is disposable. Discard after single use.
- Wipe wire set with a cleaning agent that is non-corrosive to plastic.
- Gas sterilization is permitted as needed.
- Wipe with a cleaning agent that is non-corrosive to plastic.
- Allow assembly to completely air dry before using on a patient.
- Gas sterilization is permitted as needed.
- Wipe with a cleaning agent that is non-corrosive to plastic.
- Allow assembly to completely air dry before using on a patient.
- Gas sterilization is permitted as needed.
- Wipe buckle and wire set with alcohol or a cleaning agent that is non-corrosive to plastic.
- Belt strap may be wiped with alcohol, machine washed on a gentle cycle, or soaked in disinfecting solution.
NOTE: Do not wring out RIP belts, which have a coiled wire sewn into them.
- Disassemble humidifier chamber per manufacturer’s instructions.
- Rinse chamber and brush thoroughly with a soft brush.
- Place humidifier components on the automatic washer/disinfector rack in a manner that allows each part to drain.
- Add detergent and begin the automatic washer cleaning cycle.
- Visually inspect the humidifier chamber. Discard the chamber if cracking, tears or other deterioration is apparent.
- Air dry.
- Store in a dry, dust-free environment
If manually cleaning:
- Disassemble humidifier chamber per manufacturer’s instructions.
- Rinse chamber and brush thoroughly with a soft brush using warm water.
- Soak humidifier components in an ortho-phthalaldehyde solution such as Cidex®OPA.
- Rinse thoroughly.
- Visually inspect the humidifier chamber. Discard the chamber if cracking, tears or other deterioration is apparent.
- Air dry.
- Store in a dry, dust-free environment